Table 2-11
Current range calibration commands
Step Command/procedure
1
Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
:SOUR:CURR:RANGE <Range>
:SOUR:CURR -<Source_value>
Take DMM reading.
2
:CAL:PROT:SOUR <DMM_Reading>
Check 2430 for errors.
3
:CAL:PROT:SENS <DMM_Reading>
Check 2430 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2430 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2430 for errors.
:SOUR:CURR +<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2430 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2430 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A.
Read actual output value.
Calibrate source positive zero.
2-22
Calibration
1. Perform complete procedure for each range, where <Range> and <Source_value> = 10E-6, 100E-6, 1E-3, 10E-3,
100E-3, 1, or 3.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
Step 4: Program calibration dates
Use following commands to set the calibration date and calibration due date:
:CAL:PROT:DATE <year>, <month>, <day> (Calibration date)
:CAL:PROT:NDUE <year>, <month>, <day> (Next calibration due date)
Note that the year, month, and date must be separated by commas. The allowable range for
the year is from 1998 to 2097, the month is from 1 to 12, and the date is from 1 to 31.
Step 5: Save calibration constants
Calibration is now complete, so you can store the calibration constants in EEROM by sending
the following command:
:CAL:PROT:SAVE
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......