A-6
Specifications
2430-C Specifications
Source Specifications
1
Voltage Programming Accuracy (remote sense)
Accuracy (1 Year)
Noise
Programming
23 C –5 C
(peak-peak)
Range
Resolution
–% rdg. + volts
0.1Hz
10Hz
200.000 m V
5
V
0.02% + 600
V
10
V
2.00000
V
50
V
0.02% + 600
V
50
V
20.0000
V
500
V
0.02% + 2.4 m V
500
V
100.000
V
2.5 m V
0.02% + 12 m V
2.5 m V
T E M P E R ATURE COEFFICIENT (0
18 C & 28
50 C): –(0.15
∞
accuracy specification)/ C.
MAX. OUTPUT P O W E R : 110W (four quadrant source or sink operation, DC mode).
SOURCE/SINK LIMITS: 5% over nominal for current and voltage.
V O LTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100 V.
NOISE 10Hz 1MHz (p-p): 50mV maximum, resistive load.
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = None.
C U R R E N T LIMIT:
Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
O V E R S H O O T:<0.1% typical (full scale step, resistive load, 10mA range).
Current Programming Accuracy (local or remote sense)
Accuracy (1 Year)
Noise
Programming
23 C –5 C
(peak-peak)
Range
Resolution
–% rdg. +amps
0.1Hz
10Hz
10.0000
A
500 pA
0.033% +
2 nA
50 pA
100.000
A
5 nA
0.031% +
20 nA
500 pA
1.00000 m A
50 nA
0.034% + 200 nA
5 nA
10.0000 m A
500 nA
0.045% +
2
A
50 nA
100.000 m A
5 A
0.066% +
20
A
500 nA
1.00000
A
2
50 A
0.067%
+ 400
A
100 A
3.00000
A
2
/
500 A
0.059%
+ 2.8 m A
300 A
10.0000
A
3
500 A
0.089%
+ 5.9 m A
300 A
T E M P E R ATURE COEFFICIENT (0
18 C & 28
50 C): –(0.15
∞
accuracy specification)/ C.
MAX. OUTPUT POWER: 110W (four quadrant source or sink operation, DC mode)
2
. 1100W (four quadrant source or sink operation, PULSE mode)
3
.
SOURCE/SINK LIMITS: –1.05A @ 105V, –3.15A @ 21V, –10.5A @ 105V 2.5ms pulse.
C U R R E N T R E G U L ATION: Line: 0.01% of range. Load: 0.01% of range + 100pA.
V O LTAGE LIMIT:Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
O V E R S H O O T:<0.1% typical (1mA step, R
L
= 10k , 20V range).
Section A (pgs 6-10).qxd 11/7/2005 1:35 PM Page 2
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......