![Keithley SourceMeter 2430 Скачать руководство пользователя страница 112](http://html1.mh-extra.com/html/keithley/sourcemeter-2430/sourcemeter-2430_service-manual_661803112.webp)
A-8
Specifications
Measure Specifications
1, 2, 7
Voltage Measurement Accuracy (remote sense)
Max.
Input
Accuracy (23 C – 5 C)
Range
Resolution
Resistance
1 Year, –(%rdg + volts)
200.000 m V
1 V
>10 G
0.012% + 300
V
2.00000
V
10 V
>10 G
0.012% + 300
V
20.0000
V
100 V
>10 G
0.015% +
1 m V
100.000
V
7
500 V
>10 G
0.015% +
5 m V
T E M P E R ATURE COEFFICIENT (0
18 C & 28
50 C):
–(0.15
∞
accuracy specification)/ C.
Current Measurement Accuracy (local or remote sense)
Max.
Voltage
Accuracy (23 C – 5 C)
Range
Resolution
*Burden
3
1 Year, –(%rdg + amps)
10.0000
A
100 pA
< 1mV
0.027% + 700 pA
100.000
A
1 nA
< 1mV
0.025% +
6 nA
1.00000 m A
10 nA
< 1mV
0.027% +
60 nA
10.0000 m A
100 nA
< 1mV
0.035% + 600 nA
100.000 m A
1 A
<
1mV
0.055% +
6
A
1.00000
A
10 A
<
1mV
0.060% + 120
A
3.00000
A/
30 A
<
1mV
0.052%
+1.71 m A
10.0000
A
6
30 A
<
1mV
0.082%
+1.71 m A
T E M P E R ATURE COEFFICIENT (0
18 C & 28
50 C):
–(0.1
∞
accuracy specification)/ C.
Resistance Measurement Accuracy (4 wire remote sense)
Source I Mode, Auto Ohms
Max.
Default
Normal Accuracy (23 C – 5 C)
Enhanced Accuracy (23 C – 5 C)
5
Range
Resolution
Test Current
1 Year, –(%rdg + ohms)
1 Year, –(%rdg + ohms)
<0.20000
4
1
Source I
A C C
+ Measure V
A C C
Measure I
A C C
+ Measure V
A C C
2.00000
10
1
A
0.169% + 0.0003
0.129% + 0.0001
20.0000
100
100 m A
0.098% +
0.003
0.073% +
0.001
200.000
1
m
10 m A
0.077% +
0.03
0.053% +
0.01
2.00000
k
10
m
1 m A
0.066% +
0.3
0.045% +
0.1
20.0000
k
100
m
100
A
0.063% +
3
0.043% +
1
200.000
k
1
10
A
0.065% +
30
0.046% +
10
2.00000
M
10
10
A
0.068% +
100
0.049% +
50
20.0000
M
100
1
A
0.249% +
1 k
0.112% +
500
>20.0000
M
4
1
k
Source I
A C C
+ Measure V
A C C
Measure I
A C C
+ Measure V
A C C
T E M P E R ATURE COEFFICIENT (0
18 C & 28
50 C): –(0.15
∞
accuracy specification)/ C.
SOURCE I MODE, MANUAL OHMS: Total uncertainty = I source ac V measure accuracy (4-wire remote sense).
SOURCE V MODE: Total uncertainty = V source ac I measure accuracy (4-wire remote sense).
6-WIRE OHMS MODE: Available using active ohms guard and guard sense. Max. Guard Output Current: 50 mA (except 1A and 3A
ranges). Accuracy is load dependent. Refer to manual for calculation formula.
GUARD OUTPUT IMPEDANCE: 0.1 in ohms mode.
1
Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A ranges, add 0.05%; 3A/10A range, add 1.5mA. For
0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A ranges, add 0.5%; 3A/10A range, add 15mA.
2
Accuracies apply to 2- or 4-wire mode when properly zeroed.
3
4-wire mode.
4
Manual ohms mode only.
5
Source readback enabled, offset compensation ON.
6
10A range only in pulse mode.
7
In pulse mode, limited to 0.1 PLC measurement.
Section A (pgs 6-10).qxd 11/7/2005 1:35 PM Page 4
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......