Performance Verification
3.
Configure the Model 2430 ohms function for the 4-wire sense mode as follows:
•
Press CONFIG then MEAS
Ω
. The instrument will display the following:
CONFIG OHMS
SOURCE SENSE-MODE GUARD
©
•
Select SENSE-MODE, then press ENTER. The following will be displayed:
SENSE-MODE
2-WIRE 4-WIRE
•
Select 4-WIRE, then press ENTER.
•
Press EXIT to return to normal display.
4.
Press MEAS
Ω
to select the ohms measurement function, and make sure the source out-
put is turned on.
5.
Verify ohms measurement accuracy for each of the resistance values listed in Table 1-7.
For each measurement:
•
Set the resistance calibrator output to the nominal resistance or closest available
value.
NOTE
It may not be possible to set the resistance calibrator to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
•
Select the appropriate ohms measurement range with the RANGE keys.
•
Verify that the Model 2430 resistance reading is within the limits given in the table.
6.
Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE SENSE
jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-7
Ohms measurement accuracy limits
Model 2430 range
Calibrator resistance
1
Model 2430 resistance reading limits
2
(1 year, 18°C to 28°C)
2
Ω
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
1.9
Ω
19
Ω
190
Ω
1.9k
Ω
19k
Ω
190k
Ω
1.9M
Ω
19M
Ω
1.89647 to 1.90353
Ω
18.9780 to 19.0220
Ω
189.818 to 190.182
Ω
1.89837 to 1.90163k
Ω
18.9856 to 19.0144k
Ω
189.837 to 190.163k
Ω
1.89761 to 1.90239M
Ω
18.9781 to 19.0219M
Ω
1. Nominal resistance values.
2. Reading limits based on Model 2430 normal accuracy specifications and nominal resistance values. If actual
resistance values differ from nominal values shown, recalculate reading limits using actual calibrator resistance
values and Model 2430 normal accuracy specifications. See Verification limits earlier in this section for details.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......