![Keithley SourceMeter 2430 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/keithley/sourcemeter-2430/sourcemeter-2430_service-manual_661803026.webp)
1-10
Performance Verification
2.
Select the multimeter DC volts function.
3.
Set the Model 2430 to both source and measure voltage by pressing the SOURCE V and
MEAS V keys, and make sure the source output is turned on.
4.
Verify output voltage accuracy for each of the voltages listed in Table 1-4. For each test
point:
•
Select the correct source range.
•
Set the Model 2430 output voltage to the indicated value as measured by the digital
multimeter.
•
Verify that the Model 2430 voltage reading is within the limits given in the table.
NOTE
It may not be possible to set the voltage source to the specified value. Use the closest
possible setting, and modify reading limits accordingly.
5.
Repeat the procedure for negative source voltages with the same magnitudes as those
listed in Table 1-4.
6.
Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Model 2430 source
and measure range
1
Source voltage
2
Model 2430 voltage reading
limits (1 year, 18°C to 28°C)
200mV
2V
20V
100V
200.000mV
2.00000V
20.0000V
100.000V
199.676 to 200.324mV
1.99946 to 2.00054V
19.9960 to 20.0040V
99.980 to 100.020V
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by multimeter. Use closest possible value and modify reading limits accordingly.
Output current accuracy
Follow the steps below to verify that Model 2430 output current accuracy is within specified
limits. The test involves setting the output current to each full-range value and measuring the
currents with a precision digital multimeter.
10
μ
A to 1A range accuracy
1.
With the power off, connect the digital multimeter to the Model 2430 INPUT/OUTPUT
jacks, as shown in Figure 1-2.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......