Command Reference B-5
:NDUE
(:CALibration:PROTected:NDUE)
Purpose
To send the next calibration due date to the instrument.
Format
:cal:prot:ndue <year>, <month>, <day>
Parameters
<year> = 1994 to 9999
<month> = 1 to 12
<day> = 1 to 31
Query
:cal:prot:ndue?
Response
<year>, <month>, <day>
Description
The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date by using the
:NDUE? query, or by using the front panel CAL menu.
Note
The next due date parameters must be delimited by commas.
Example
:CAL:PROT:NDUE 1999,11,20
Send due date (11/20/99).
:SENSE
(:CALibration:PROTected:SENSe)
Purpose
To calibrate the sense function.
Format
:cal:prot:sens <nrf>
Parameters
See Table B-2 for recommended parameter ranges for optimum calibration.
Description
The :CAL:PROT:SENS command calibrates the Model 2430 sense function.
To use this command, you must (1) program the source to the correct value,
(2) select the range being calibrated, and (3) send the :CAL:PROT:SENS
command for each parameter listed in Table B-2.
Note
When the Model 2430 receives this command, it will attempt to place the pa-
rameter into one of three parameter ranges summarized in Table B-2, de-
pending on the active :SENS:FUNC and range. If the parameter does not fit
into any of the three allowed ranges, an error number -222, “Parameter data
out of range” will be generated. Once the unit has successfully selected the
appropriate parameter range, it will then check to see if autorange for the ac-
tive :SENS:FUNC is enabled. If so, an error number -221, “Settings conflict”
will be generated. If no error occurs, the active sense function/range point
will be calibrated using the corresponding parameter.
Note that parameters for a given sense function and range may be sent in any
order; however, once one :CAL:PROT:SENS command executes, the other
two must also be executed, or an error number -200, “Execution error” will
occur when the :CAL:PROT:SAVE command is processed at the end of the
calibration procedure.
Example
:CAL:PROT:SENS 2
Calibrate 2V sense range.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......