B-6
Command Reference
Table B-2
Recommended :CALibration:PROTected:SENSe parameter ranges
Sense
range
First parameter
(zero)
Second parameter
(negative full scale)
Third parameter
(positive full scale)
0.2V
2V
20V
100V
0.002 to +0.002
0.02 to +0.02
0.2 to +0.2
1 to +1
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-90 to -110
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+90 to +110
10
μ
A
100
μ
A
1mA
10mA
100mA
1A
3A
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to 3E-2
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -11E-3
-90E-3 to -110E-3
-0.9 to -1.1
-2.7 to -3.15
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+2.7 to +3.15
Note: Actual allowable parameter ranges are ±50% of full range for zero parameters, 75% of full
scale to 150% of full scale for full-range parameters, but ranges listed above should be observed
for optimum calibration.
:SOURCE
(:CALibration:PROTected:SOURce)
Purpose
To calibrate the source function.
Format
:cal:prot:sour <nrf>
Parameters
See Table B-3 for recommended parameter ranges.
Description
The :CAL:PROT:SOUR command calibrates the Model 2430 source func-
tion. To use this command, you must (1) program the source to the correct
value, (2) select the range being calibrated, and (3) send the
:CAL:PROT:SOUR command for each parameter listed in Table B-3.
Note
When the Model 2430 receives this command, it will attempt to place the pa-
rameter into one of the four parameter ranges summarized in Table B-3, de-
pending on the active :SOUR:FUNC and range. If the parameter does not fit
into any of the four allowed ranges, an error number -222, “Parameter data
out of range” will be generated. Once the unit has successfully selected the
appropriate parameter range, it will then check to see if the active source is
programmed to a value within the selected parameter range, and that
:OUTP:STAT is ON. If the active source is not programmed to a value with-
in the parameter range (or if the source is in autorange or in standby), an error
number -221, “Settings conflict” will be generated. If no error occurs, the ac-
tive source function/range point will be calibrated using the corresponding
parameter.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......