![Keithley SourceMeter 2430 Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/keithley/sourcemeter-2430/sourcemeter-2430_service-manual_661803135.webp)
Calibration Programs C-7
Requesting calibration constants
Program C-2 listed below demonstrates how to request the calibration constants from the
Model 2430. The program requests and displays the calibration constants for all ranges of both
the volts and current source and measurement functions.
Program C-2 Requesting calibration constants
' Example program to request calibration constants.
' Model 2430 primary address = 24
OPEN "IEEE" FOR OUTPUT AS #1 ' Open IEEE-488 output path.
OPEN "IEEE" FOR INPUT AS #2 ' Open IEEE-488 input path.
PRINT #1, "INTERM CRLF" ' Set input terminator.
PRINT #1, "OUTTERM LF" ' Set output terminator.
PRINT #1, "REMOTE 24" ' Put 2430 in remote.
PRINT #1, "OUTPUT 24;*RST" ' Restore defaults.
CLS
PRINT #1, "OUTPUT 24;:SENS:FUNC:CONC OFF"
PRINT #1, "OUTPUT 24;:SENS:FUNC 'VOLT:DC'"
PRINT #1, "OUTPUT 24;:SOUR:FUNC VOLT"
PRINT #1, "OUTPUT 24;:CAL:PROT:CODE 'KI002430'" ' Unlock calibration.
Range = .2 ' Initial range = 200mV.
PRINT "Model 2430 Voltage Calibration Constants:": PRINT
DO ' Loop for all volts ranges.
PRINT #1, "OUTPUT 24;:SOUR:VOLT:RANGE "; Range
PRINT #1, "OUTPUT 24;:CAL:PROT:SENS:DATA?"
PRINT #1, "ENTER 24"
LINE INPUT #2, Data$
PRINT Data$
PRINT #1, "OUTPUT 24;:CAL:PROT:SOUR:DATA?"
PRINT #1, "ENTER 24"
LINE INPUT #2, Data$
PRINT Data$
Range = Range * 10
IF Range = 200 THEN Range = 100
LOOP WHILE Range <= 100
PRINT #1, "OUTPUT 24;:CAL:PROT:LOCK"
PRINT #1, "OUTPUT 24;:SENS:FUNC 'CURR:DC'"
PRINT #1, "OUTPUT 24;:SOUR:FUNC CURR"
PRINT #1, "OUTPUT 24;:CAL:PROT:CODE 'KI002430'"
Range = .00001 ' Initial range = 10uA.
PRINT : PRINT "Model 2430 Current Calibration Constants:": PRINT
DO ' Loop for all current ranges.
PRINT #1, "OUTPUT 24;:SOUR:CURR:RANGE "; Range
PRINT #1, "OUTPUT 24;:CAL:PROT:SENS:DATA?"
PRINT #1, "ENTER 24"
LINE INPUT #2, Data$
PRINT Data$
PRINT #1, "OUTPUT 24;:CAL:PROT:SOUR:DATA?"
PRINT #1, "ENTER 24"
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......