B-2
Command Reference
Introduction
This appendix contains detailed information on the various Model 2430 remote calibration
commands, calibration error messages, and methods to detect the end of each calibration step.
Section 2 of this manual covers detailed calibration procedures.
Command summary
Table B-1 summarizes Model 2430 calibration commands. These commands are covered in
detail in the following paragraphs.
Table B-1
Remote calibration command summary
Command
Description
:CALibration
:PROTected
:CODE '<password>'
:COUNT?
:SAVE
:LOCK
:LOCK?
:DATE <y>,<m>,<d>
:DATE?
:NDUE <y>,<m>,<d>
:NDUE?
:SENSe <nrf>
:DATA?
:SOURce <nrf>
:DATA?
Calibration subsystem.
Cal commands protected by password.
Unlock cal; if cal is unlocked, changes password.
(Default password: KI002430.)
Query number of times 2430 has been calibrated.
Save calibration data to EEPROM.*
Lock calibration, inhibit SAVE command operation.
Request cal lock status. (0 = unlocked; 1= locked.)
Program calibration year, month, day.
Query calibration year, month, day.
Program calibration due year, month, day.
Query calibration due year, month, day.
Calibrate active measure range.
Query measurement cal constants for active range.
Calibrate active source range.
Query source cal constants for active range.
* Calibration data will not be saved if:
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed. (For example, omitting a negative
full-scale step.) Ranges that were successfully calibrated will be saved if calibration is unlocked.
Ranges that failed will not be saved.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......