![Keithley SourceMeter 2430 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/keithley/sourcemeter-2430/sourcemeter-2430_service-manual_661803047.webp)
Calibration 2-15
3.
Select SAVE, then press ENTER. The unit will display the following message:
SAVE CAL
Press ENTER to continue; EXIT to abort calibration sequence
4.
Press ENTER. The unit will prompt you for the calibration date:
CAL DATE: 12/15/1998
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
5.
Change the displayed date to today's date, then press the ENTER key. Press ENTER
again to confirm the date.
6.
The unit will then prompt for the calibration due date:
NEXT CAL: 12/15/1998
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
7.
Set the calibration due date to the desired value, then press ENTER. Press ENTER again
to confirm the date.
8.
Once the calibration dates are entered, calibration is complete, and the following mes-
sage will be displayed:
CALIBRATION COMPLETE
Press ENTER to confirm; EXIT to abort
9.
Press ENTER to save the calibration data (or press EXIT to abort without saving cali-
bration data.) The following message will be displayed:
CALIBRATION SUCCESS
Press ENTER or EXIT to continue.
10.
Press ENTER or EXIT to complete process.
Step 5: Lock out calibration
1.
From normal display, press MENU.
2.
Select CAL, then press ENTER. The Model 2430 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
3.
Select LOCK, then press ENTER to lock out calibration.
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands over
the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are sepa-
rately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
these commands, refer to Appendix B.
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......