Specifications
A-7
Additional Source Specifications
TRANSIENT RESPONSE TIME: 30 s typical for the output to recover to its spec. following a step change in load. Resistive load.
COMMAND PROCESSING TIME: Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent
<nrf> command.Autorange On: 10ms. Autorange Off
: 7ms.
O U T P U T SETTLING TIME: Time required to reach 0.1% of final value after command is processed. 100 s typical. Resistive load. 10 A to 100mA
range, 1V step.
O U T P U T S L E W R ATE: 0.25V/ s –30%, 100V range, 100mA compliance. 0.08V/ s –30%, 20V range, 100mA compliance.
C O M M O N M O D E V O LTAGE: 250VDC.
REMOTE SENSE: Up to 1V drop per load lead.
COMPLIANCE A C C U R A C Y:For Source mode, add 0.1% of range to base specification. For Sink mode, 10 A to 100mA range, accuracy is –(0.5%
+ offset*3). For Sink mode, 1A and 3A ranges, accuracy is –(1.5% + offset*3).
OVER T E M P E R ATURE PROTECTION: Internally sensed temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT:Overshoot into a fully resistive 100k load, 10Hz to 1MHz BW, adjacent ranges: (100mV) typical, except 20V/100V
range boundary.
MINIMUM COMPLIANCE VALUE: 0.1% of range.
C O N TACT CHECK:
2ž
15ž
50ž
No contact check failure
<1.00
<13.5
<47.5
Always contact check failure
>3.00
>16.5
>52.5
Additional Pulse Mode Source Specifications
MAXIMUM DUTY CYCLE: 8%, hardware limited, 10A range only.
All other ranges 84%.
MAXIMUM PULSE WIDTH: 5ms from 90% rising to 90% falling edge, 2.5ms 10A range.
MINIMUM PULSE WIDTH: 150 s, with 0.004 PLC measurement.
MINIMUM PULSE RESOLUTION: 10 s typical, 20 s max., limited by system jitter.
SOURCE A C C U R A C Y:Determined by settling time and source range specifications.
O U T P U T SETTLING TIME, 0.1%:
800 s typ., source I = 10A into 10 , limited by voltage slew rate.
500 s typ., source I = 10A into 1 , limited by voltage slew rate.
O U T P U T SLEW RATE:Voltage (10 load): 0.25V/ s –30% on 100V range
0.08V/ s –30% on 20V range, 10A range.
Current (0 load):
0.25A/ s –30% on 100V range
0.08A/ s –30% on 20V range, 10A range.
1
Specifications valid for continuous output currents below 105mA. For operating above 105mA on the 1A range for >1 minute, derate accuracy 10%/100mA
above 105mA. For operating above 105mA on the 3A range for >1 minute, derate accuracy 10%/300mA above 105mA.
2
Full power source operation regardless of load to 30 C. Above 30 C and/or power sink operation, refer to the Power Equations section of the User s Manual.
For power line voltages less than 95V, limit max. current to 0.85A on 1A range and 2.5A on 3A range.
3
10A range only in pulse mode. Limited to 2.5ms pulse width maximum, 8% duty cycle maximum.
Section A (pgs 6-10).qxd 11/7/2005 1:35 PM Page 3
Содержание SourceMeter 2430
Страница 1: ...Contains Calibrating and Servicing Information Model 2430 1kW Pulse SourceMeter Service Manual ...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Страница 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Страница 16: ......
Страница 17: ...1 Performance Verification ...
Страница 33: ...2 Calibration ...
Страница 56: ...2 24 Calibration ...
Страница 57: ...3 Routine Maintenance ...
Страница 60: ...3 4 Routine Maintenance ...
Страница 61: ...4 Troubleshooting ...
Страница 75: ...5 Disassembly ...
Страница 82: ...5 8 Disassembly ...
Страница 83: ...6 Replaceable Parts ...
Страница 104: ...6 22 Replaceable Parts ...
Страница 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Страница 116: ...A 12 Specifications ...
Страница 117: ...B Command Reference ...
Страница 128: ...B 12 Command Reference ...
Страница 129: ...C Calibration Programs ...
Страница 140: ......