Model 2110 5½ Digit Multimeter Reference Manual
Section 5: Remote commands
2110-901-01 Rev. C/August 2013
5-29
CALCulate:FUNCtion <name>
This command configures the math function for the calculate commands. You can only select one function at a
time.
Parameters
<name>
Description
PERCent
Select percent math calculation.
AVERage
Select average math calculation.
NULL
Select the null feature.
LIMit
Select the limits feature to configure and control limits testing.
MXB
Select mX+b polynomial math calculation.
DB
Select the mX+b calculations.
DBM
Select the dBM reference resistance value.
Query
command
description
FUNCtion?
Query the measurement function.
Details
Refer to
(on page 3-48) for detail about the math operations.
CALCulate:LIMit:LOWer <n>|MIN|MAX
This command is used to specify the lower limit for limit testing.
A limit value of 1 is 1 V for the voltage functions (DCV or ACV), 1 A for the current functions (DCI or ACI), 1
Ω
on
the resistance functions (2 or 4), and 1 (C, F, or K) for the temperature function. A limit value is not range
sensitive. For example, a limit of 1 for DCV is 1 V on all measurement ranges.
Parameter
<n>
description
0 to ± 120% of the
highest range
Set the lower limit for limit testing. You can set the value to any number from 0
to ± 120% of the highest range, for the present function.
MINimum
Set specified limit to 0.
MAXimum
Set specified limit to ± 120% of the highest range.
Query
command
description
CALCulate:LIMit:LOWer?
Query the lower limit for the limit testing.
CALCulate:LIMit? MIN
Query the minimum value.
CALCulate:LIMit? MAX
Query the maximum value.