![Keithley 2110 Скачать руководство пользователя страница 205](http://html.mh-extra.com/html/keithley/2110/2110_reference-manual_661855205.webp)
Appendix C: Communication interface
Model 2110 5½ Digit Multimeter Reference Manual
C-6
2110-901-01 Rev. C/August 2013
Figure 48: USB connector
Contact number
Signal name
Typical wiring assignment
Description
1
VBUS
Red
Floating
2
D-
White
Limit Test Pass
3
D+
Green
Limit Test Fail
5
GND
Black
GND
If you turn off the USB interface, the Pass/Fail output function will turn on automatically.
If you want to use the Pass/Fail signal output, you must use the GPIB interface for remote control.
Disconnect the USB cable from your instrument while using the Pass/Fail signal output.
To turn on and off the pass/fail output:
1. Press the
SHIFT
key, and then press the
CONFIG
key.
2.
Use the ► and ◄ keys to select
INTERFACE
.
3. Press
ENTER
.
4.
Use the ► and ◄ keys to:
•
Select
GPIB
to turn on the output.
•
Select
USB
to turn off the output.
5. Press
ENTER
.