Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-36
2110-901-01 Rev. C/August 2013
[SENSe:]CAPacitance:RANGe <n>|MIN|MAX
Use this command to select the capacitance range.
Parameters
<n>
description
0 to 10e-3
Specify the value of the expected capacitance reading.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command
description
[SENSe:]CAPacitance:RANGe
Query the capacitance range.
[SENSe:]CAPacitance:RANGe? MIN
Query the minimum capacitance range.
[SENSe:]CAPacitance:RANGe? MAX
Query the maximum capacitance range.
Details
The measurement ranges for capacitance measurements are 1 nF, 1
0 nF, 100 nF, 1 μF, 10 μF, 100 μF, 1 mF,
and 10 mF.
Resolution is ignored for capacitance. Refer to
(on page 3-24) for more information
about this command.
[SENSe:]CAPacitance:RANGe:AUTO <b>
This command turns capacitance autorange on or off.
Parameters
<b>
description
0 or OFF
Turn autorange off.
1 or ON
Turn autorange on.
Query
command
description
[SENSe:]CAPacitance:RANGe:AUTO?
Query autorange state.