Model 2110 5½ Digit Multimeter Reference Manual
Section 5: Remote commands
2110-901-01 Rev. C/August 2013
5-49
[SENSe:]RESistance:NPLCycles <n>|MIN|MAX
Use this command to select the integration time for 2-wire resistance measurements. Integration time for 2-wire
measurements is expressed as the number of power line cycles (NPLC).
Parameters
<n>
description
0.001
Select 0.001 NPLC.
0.006
Select 0.006 NPLC.
0.02
Select 0.02 NPLC.
0.06
Select 0.06 NPLC.
0.2
Select 0.2 NPLC.
0.6
Select 0.6 NPLC.
1
Select 1 NPLC.
2
Select 2 NPLC.
10
Select 10 NPLC.
100
Select 100 NPLC.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command
description
[SENSe:]RESistance:NPLCycles?
Query the NPLCycles.
[SENSe:]RESistance:NPLCycles? MIN
Query the minimum value.
[SENSe:]RESistance:NPLCycles? MAX
Query the maximum value.
Details
Refer to
Integration time and resolution for DC and resistance measurements
(on page 3-41) more information
about resolution.
[SENSe:]TCOuple:RJUNction:REAL?
Use this command to query the temperature of the internal reference junction.
Query
command
description
[SENSe:]TCOuple:RJUNction:REAL?
Query the temperature of the internal reference
junction.
Details
The junction temperature is reported in degrees Celsius.
Refer to the
(on page 3-17) section for more information about offset value.