Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-44
2110-901-01 Rev. C/August 2013
[SENSe:]FREQuency:VOLTage:RANGe <n>|MIN|MAX
Use this command to select the voltage range for the frequency function.
Parameters
<n>
description
0 to 750
Set AC voltage range for frequency function.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command
description
[SENSe:]FREQuency:VOLTage:RANGe?
Query the measurement range.
[SENSe:]FREQuency:VOLTage:RANGe? MIN
Query the minimum range.
[SENSe:]FREQuency:VOLTage:RANGe? MAX
Query the maximum range.
[SENSe:]FREQuency:VOLTage:RANGe:AUTO <b>
This command turns autorange on or off.
Parameters
<b>
description
0 or OFF
Turn autorange off.
1 or ON
Turn autorange on.
Query
command
description
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?
Query autorange state.