Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-22
2110-901-01 Rev. C/August 2013
Command
Description
:NPLCycles <n>|MIN|MAX
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.06, 1, 2, 10, 100,
MIN, or MAX. Refer to
Integration time and resolution for DC and
(on page 3-41) for more information
about valid values for n.
:NPLCycles?[MIN|MAX]
Query NPLCycles.
:VOLTage:AC
Path to configure AC voltage.
:RANGe <n>|MIN|MAX
Specify range: 100 mV, 1 V, 10 V, 100 V, 750 V, MIN, or MAX.
? [MIN|MAX]
Query the range.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution. Where <n> = 1.00e-06 to 1.00e-04 of range,
MIN, or MAX. Refer to
Bandwidth and resolution for AC
(on page 3-44) for more information about valid
values for n.
:RESolution?
Query the resolution.
:CURRent:AC
Path to configure AC current.
:RANGe <n>|MIN|MAX
Specify range:0 to 10, MIN, or MAX.
? [MIN|MAX]
Query the range for AC voltage.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution. Where <n> = 1.00e-06 to 1.00e-04 of range,
MIN, or MAX. Refer to
Bandwidth and resolution for AC
(on page 3-44) for more information about valid
values for n.
:RESolution? [MIN|MAX]
Query the resolution.
:CURRent[:DC]
Path to configure DC current.
:RANGe <n>|MIN|MAX
Specify range: 10 mA, 100 mA, 1 A, 3 A, 10 A, MIN, or MAX.
? [MIN|MAX]
Query the range for DC voltage.
:AUTO <b>
Enable or disable auto range. Where: 0 = off and 1 = on.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to
Integration time and resolution for DC and
(on page 3-41) for more information
about valid values for n.
:RESolution ?
Query the resolution.
:NPLCycles <n> MIN|MAX
Specify the integration time in number of power line cycles for
the selected function: 0.001, 0.006, 0.02, 0.06, 1, 2, 10, 100,
MIN, or MAX. Refer to
Integration time and resolution for DC and
(on page 3-41) for more information
about valid values for n.
:NPLCycles? [MIN|MAX]
Query NPLCycles.
:RESistance
Path to configure resistance (2-wire
Ω
).
:RANGe <n>|MIN|MAX
Specify range: 0 to 100e6, MIN, or MAX.
? [MIN|MAX]
Query resistance range.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to
Integration time and resolution for DC and
(on page 3-41) for more information
about valid values for n.
:RESolution? [MIN|MAX]
Query the resistance resolution.