Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-28
2110-901-01 Rev. C/August 2013
CALCulate:DB:REFerence <n>|MIN|MAX
This command stores a relative value in the dB Relative Register. You must turn on the math operation before
writing to the math register.
Parameters
parameters
description
<n>
Set the dB value to any value between -200 and 200.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command
description
CALCulate:DB:REFerence?
Query the dB relative value.
CALCulate:DB:REFerence? MIN
Query the minimum value.
CALCulate:DB:REFerence? MAX
Query the maximum value.
CALCulate:DBM:REFerence <n>|MIN|MAX
This command is used to set the dBm reference resistance value in ohms.
Parameters
parameter
description
<n>
You can set the relative value to any number between 50 ~ 8000
Ω
.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command
description
CALCulate:DBM:REFerence?
Query the dBm reference value.
CALCulate:DBM:REFerence? MIN
Query the minimum value.
CALCulate:DBM:REFerence? MAX
Query the maximum value.