Section 3: Basic DMM operation
Model 2110 5½ Digit Multimeter Reference Manual
3-40
2110-901-01 Rev. C/August 2013
Remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
Use the following commands to set a measurement range:
<function>:RANGe <n>
Specify expected reading
<function>
=
VOLTage[:DC]
<n>
= -1000 to 1000
Volts DC
=
VOLTage:AC
= 0 to 750
Volts AC
=
CURRent[:DC]
= -10 to 10
Amps DC
=
CURRent:AC
= 0 to 10
Amps AC
=
RESistance
= 0 to 100e6
Ohms (2-wire)
=
FRESistance
= 0 to 100e6
Ohms (4-wire)
=
FREQuency[:VOLT]
= 0 to 750
Volts AC (voltage input)
=
FREQuency:CURRent
= 0 to 10
Amps AC (current input)
=
PERiod[:VOLT]
= 0 to 750
Volts AC (voltage input)
=
PERiod:CURRent
= 0 to 10
Amps AC (current input)
=
CAPacitance
= 0 to 10e-3
Farads
<function>:RANGe:AUTO <b>
Enable or disable autorange
<function>
=
VOLTage[:DC]
<b>
= 0 or OFF
Disable autorange
=
VOLTage:AC
= 1 or ON
Enable autorange
=
CURRent[:DC]
=
CURRent:AC
=
RESistance
=
FRESistance
=
FREQuency[:VOLT]
=
FREQuency:CURRent
=
PERiod[:VOLT]
=
PERiod:CURRent
=
CAPacitance