Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-38
2110-901-01 Rev. C/August 2013
[SENSe:]CURRent[:DC]:NPLCycles <n>|MIN|MAX
Use this command to select the integration time for DC current measurements. Integration time for DCI
measurements is expressed as the number of power line cycles (NPLC).
Parameters
<n>
description
0.001
Select 0.001 NPLC.
0.006
Select 0.006 NPLC.
0.02
Select 0.02 NPLC.
0.06
Select 0.06 NPLC.
0.2
Select 0.2 NPLC.
0.6
Select 0.6 NPLC.
1
Select 1 NPLC.
2
Select 2 NPLC.
10
Select 10 NPLC.
100
Select 100 NPLC.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command
description
[SENSe:]CURRent[:DC]:NPLCycles?
Query the NPLCycles.
[SENSe:]CURRent[:DC]:NPLCycles? MIN
Query the minimum value.
[SENSe:]CURRent[:DC]:NPLCycles? MAX
Query the maximum value.
Details
Refer to
Integration time and resolution for DC and resistance measurements
(on page 3-41) for more
information about NPLC.
[SENSe:]CURRent[:DC]:RANGe <n>|MIN|MAX
Use this command to select the function (DCI) and set a measure range.
Parameters
<n>
description
-10 to 10
Select the measurement range in amps.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
Query
command
description
[SENSe:]CURRent[:DC]:RANGe?
Query the measurement range.
[SENSe:]CURRent[:DC]:RANGe? MIN
Query the minimum range.
[SENSe:]CURRent[:DC]:RANGe? MAX
Query the maximum range.