Model 2110 5½ Digit Multimeter Reference Manual
Section 5: Remote commands
2110-901-01 Rev. C/August 2013
5-21
DISPlay subsystem summary
Command
Description
:DISPlay <b>
Turn on or off front panel display. Where: 0 = off and 1 = on.
:DISPlay:TEXT <a>
Define ASCII message “a” (up to 16 characters in length). The message
will display on the secondary display.
:DISPlay:TEXT:CLEar
Clear the message shown on the front panel display.
:DISPlay?
Query the display setting. Returns 0 (off) or 1 (on).
SENSe subsystem summary
See the following table for the SENSe command summary. Refer to
(on page 1-6)
for information about default values for SENSe commands.
Command
Description
[:SENSe]
Path to sense commands.
:FUNCtion[1|2]
Path to FUNCtion, FUNCtion1, and FUNCtion2. FUNCtion1
displays the selected function on the primary display. FUNCtion2
displays the selected function on the secondary display.
<a>
Specify one of the following functions to measure (enclosed
function in quotation marks):
"VOLTage:AC"
"VOLTage[:DC]"
"VOLTage[:DC]:RATio"
"CURRent:AC"
"CURRent[:DC]"
"FREQuency[:VOLT]"
"FREQuency:CURR"
"FRESistance"
"PERiod[:VOLT]"
"PERiod:CURR"
"RESistance"
"DIODe"
"TCOuple"
"TEMPerature"
"CONTinuity"
Note: DIODe and CONTinuity are not SCPI compliant
parameters.
:FUNCtion[1|2]?
Query function. Returns function name.
:VOLTage[:DC]
Path to configure DC voltage.
:RANGe <n>|MIN|MAX
Specify the range: 100 mV, 1 V, 10 V, 100 V, 1000 V, MIN, or
MAX.
? [MIN|MAX]
Query the range.
:AUTO <b>
Enable or disable auto range. Where: 0 = OFF and 1 = ON.
:AUTO?
Query auto range.
:RESolution <n>|MIN|MAX
Specify resolution: <n> = 3.00e-07 to 3.00e-04 of range, MIN, or
MAX. Refer to
Integration time and resolution for DC and
(on page 3-41) for more information
about valid values for n.
:RESolution? [MIN|MAX]