Section 3: Basic DMM operation
Model 2110 5½ Digit Multimeter Reference Manual
3-42
2110-901-01 Rev. C/August 2013
Integration time and resolution settings for DCV
Display resolution
Integration
time
Measurement resolution (remote
operation)
4½ digits
0.001 PLC
0.0003 x full scale range
4½ digits
0.006 PLC
0.0002 x full scale range
4½ digits
0.02 PLC
0.0001 x full scale range
4½ digits
0.06 PLC
0.00005 x full scale range
5½ digits
0.2 PLC
0.00001 x full scale range
5½ digits
0.6 PLC
0.000005 x full scale range
5½ digits
1 PLC
0.000003 x full scale range
5½ digits
2 PLC
0.000002 x full scale range
5½ digits
10 PLC
0.000001 x full scale range
5½ digits
100 PLC
0.0000003 x full scale range
When you enable math operations and your integration rate is faster than 0.6 PLC, the maximum
reading rate is affected significantly. For example, the maximum rate at 0.001 PLC with a math
function enabled is 6000 readings per second. For more detail, see
(on page 3-48).
Front-panel operation
Perform the following steps to set display resolution:
1. Select the DCV, DCI,
Ω
2, or
Ω
4 measurement function.
2. Use the
DIGITS
key to set display resolution. Each key-press toggles the display between 4½
and 5½ digits.
Perform the following steps to set the integration time (speed):
1. Select the DCV, DCI,
Ω
2, or
Ω
4 measurement function.
2. Press the
CONFIG
key, use the ◄ and ► keys to display NPLC, and press
ENTER
.
3.
Use the ◄ and ► keys to display a PLC value and press
ENTER
.