![Keithley 2110 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/keithley/2110/2110_reference-manual_661855060.webp)
Model 2110 5½ Digit Multimeter Reference Manual
Section 3: Basic DMM operation
2110-901-01 Rev. C/August 2013
3-17
Temperature measurement
For temperature measurements, the instrument supports resistance temperature detector (RTD)
probes, NTC thermistors, and thermocouple sensors.
In general, RTDs have a better accuracy and long-term stability than thermocouples. For the
instrument, the default RTD sensor type is PT100. The default thermocouple sensor is type K.
RTD and thermistor measurements
The instrument can perform temperature measurements using resistance temperature detectors
(RTDs). The supported RTD sensor types are PT100, D100, F100, PT385, PT3916, user definable,
and SPRTD. The supported thermistor type is NTCT. The instrument supports 2-wire, 3-wire, and 4-
wire RTD probes.
The default RTD sensor type is PT100. The temperature measurement range for the PT100 sensor is
-200
°
C to 850
°
C (-328
°
F to 1562
°
F).
Front-panel operation
Perform the following steps to measure temperature using an RTD or thermistor:
1. Referring to the drawing below, connect the RTD (or thermistor) to the Model 2110.
2. Press the
TEMP
and
CONFIG
keys.
3. Select the sensor:
a. Use the ◄ or ► key to display
SENSOR
and press
ENTER
.
b. Use the ◄ and ► keys to display the sensor and press
ENTER
.
Note: If you select the
NTCT
,
SPRTD
or
USER
sensor, you will be need to enter coefficient
(on page 3-20) for details.
4. Select the measurement units:
a. Press
CONFIG
, use the ► key to display
UNITS
, and then press
ENTER
.
b. Use the ◄ or ► key to display the units (
°C
,
°F
, or
K
) and press
ENTER
.
5. Select the transducer type (2-wire or 4-wire):
a. Press
CONFIG
, use the ► key to display
TRANSDUCER
, and then press
ENTER
.
b. Use the ◄ or ► key to display the transducer type (
2W RTD
or
4W RTD
), and press
ENTER
.
6. Observe the temperature reading on the display.