Model 2110 5½ Digit Multimeter Reference Manual
Section 5: Remote commands
2110-901-01 Rev. C/August 2013
5-31
CALCulate:MXB:MMFactor <n>|MIN|MAX
This command defines the m factor for the mX+b calculation.
Parameters
<n>
description
-1e6 to 1e6
Set a value for m.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command
description
CALCulate:MXB:MMFactor?
Query the value of m.
CALCulate:MXB:MMFactor? MIN
Query the minimum value of m.
CALCulate:MXB:MMFactor? MAX
Query the maximum value of m.
CALCulate:NULL:OFFSet <n>|MIN|MAX
This command stores a null value in the instrument’s Null Register. You must turn on the math operation before
writing to the math register.
Parameters
<n>
description
0 to ± 120% of the
highest range
Select the null (relative) value. You can set the null value to any number
between -120 % and 120 % of the highest range.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command
description
CALCulate:NULL:OFFSet?
Query the null value.
CALCulate:NULL:OFFS? MIN
Query the null minimum value for null math function.
CALCulate:NULL:OFFS? MAX
Query the null maximum value for null math function.
CALCulate:PERCent:TARGet <n>|MIN|MAX
This command specifies the target value for percent math function.
Parameters
<n>
description
Specify the measurement function dependent target value.