![Keithley 2110 Скачать руководство пользователя страница 172](http://html.mh-extra.com/html/keithley/2110/2110_reference-manual_661855172.webp)
Model 2110 5½ Digit Multimeter Reference Manual
Section 5: Remote commands
2110-901-01 Rev. C/August 2013
5-65
SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in
(on page 5-25).
:BEEPer <b>
This command turns the beeper on or off for limit tests, diode tests, and reading hold.
Parameters
<b>
description
0 or OFF
Turn beeper off.
1 or ON
Turn beeper on.
Query
command
description
BEEPer?
Query the state of the beeper. Returns 0 (off) or 1 (on).
Example
BEEPer ON
Turns on the beeper.
:DATA:POINts?
Query the number of readings stored in the instrument’s internal memory.
Parameters
None.
Query
command
description
:DATA:POINTs?
Query the number of readings stored in the multimeter’s internal memory.
SYSTem:ERRor?
Query the multimeter’s error queue. Up to 20 errors can be stored in the queue. Errors are retrieved in first-in-first-
out (FIFO) order. Each error string may contain up to 80 characters.
Query
This query command returns an error code. For example:
-102 Syntax error
Refer to
(on page 6-3) for information about the error codes.