Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-68
2110-901-01 Rev. C/August 2013
TRIGger:COUNt <n>|MIN|MAX|INFinite
Use this command to specify the number of triggers the instrument will accept before returning to the “idle” state.
Select from 1 to 50,000 triggers. Trigger count is ignored while in local operation.
Parameters
<n>
description
1 to 50,000
Select value for the number of triggers.
MINimum
Select the minimum value.
MAXimum
Select the maximum value.
INFinite
Select INFinite to instruct the instrument to take triggers continuously.
Query
command
description
TRIGger:COUNt?
Query the number of triggers. Returns “9.9037” when you
select the infinite trigger count.
TRIGger:COUNt? MIN
Query the minimum number of triggers.
TRIGger:COUNt? MAX
Query the maximum number of triggers.
TRIGger:DELay <n>|MIN|MAX
Use this command to select a trigger delay time in seconds. The delay is the time between the trigger signal and
each sample that follows. Specify a delay time from 0 to 3600 seconds.
Parameters
<n>
description
0 to 3600
Specify a delay time from 0 to 3600 seconds.
Query
command
description
TRIGger:DELay?
Query the trigger delay time.
TRIGger:DELay? MAX
Query the minimum trigger delay time.
TRIGger:DELay? MIN
Query the maximum trigger delay time.
Details
Refer to
(on page 3-34) for information about sources.