Section 5: Remote commands
Model 2110 5½ Digit Multimeter Reference Manual
5-30
2110-901-01 Rev. C/August 2013
CALCulate:LIMit:UPPer <n>|MIN|MAX
This command is used to specify the upper limit for limit testing. The actual limit depends on which measurement
function is currently selected.
For example, a limit value of 1 is 1 V for the volts functions (DCV or ACV), 1 A for the current functions (DCI or
ACI), 1
Ω
on the ohms functions (2-wire or 4-wire), and 1 (C, F, or K) for the temperature function. A limit value is
not range sensitive. A limit of 1 for DCV is 1 V on all measurement ranges.
Parameters
<n>
description
0 to ± 120% of the
highest range
Set the upper limit for limit testing. You can set the value to any number
from 0 to ± 120% of the highest range, for the present function.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command
description
CALCulate:LIMit:UPPer?
Query the upper limit for the limit testing.
CALCulate:LIMit:UPPer? MIN
Query the minimum upper limit.
CALCulate:LIMit:UPPer? MAX
Query the maximum upper limit.
CALCulate:MXB:MBFactor <n>|MIN|MAX
This command defines the b factor for the mX+b calculation.
Parameters
<n>
description
-1e6 to 1e6
Set a value for b.
MINimum
Set value to the minimum.
MAXimum
Set value to the maximum.
Query
command
description
CALCulate:MXB:MBFactor?
Query the value of b.
CALCulate:MXB:MBFactor? MIN
Query the minimum value of b.
CALCulate:MXB:MBFactor? MAX
Query the maximum value of b.