Section 3: Basic DMM operation
Model 2110 5½ Digit Multimeter Reference Manual
3-54
2110-901-01 Rev. C/August 2013
Remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
Use the following commands to configure and control limit testing:
CALCulate:FUNCtion LIMit
Select the limits feature
CALCulate:LIMit:LOWer <n>
Set the lower limit
CALCulate:LIMit:UPPer <n>
Set the upper limit
<n>
= (see Note)
Set value (1
µ
resolution)
The limit value setting range depends on the selected function. For example, for DCV, the limit value
can be set from -1200 to 1200.
CALCulate:STATe <b>
Enable or disable math operations
<b>
= 0 or OFF
Disable math
= 1 or ON
Enable math
mX+b
This math function multiplies a scale factor (m) to the measured reading (X), and then adds an offset
(b). The result of the calculation (y) is displayed on the instrument:
y = mX + b
You can enable this math function and set the values of m and b from the front panel or by using
remote operation. These values are stored in volatile memory and are lost when the instrument is
turned off.