Section 3: Basic DMM operation
Model 2110 5½ Digit Multimeter Reference Manual
3-20
2110-901-01 Rev. C/August 2013
Entering sensor coefficients
There are three RTD sensor types that allow you to enter coefficient values for sensor parameters.
The coefficients are provided with the documentation supplied by the manufacturer of the RTD.
The following sensors allow you to enter coefficient values:
•
USER sensor
•
NTCT sensor
•
SPRTD sensor
Perform the following steps to enter coefficient values:
1. Press the
TEMP
key.
2. Press
CONFIG
, and with
SENSOR
displayed, press
ENTER
.
3.
Use the ◄ or ► key to display the sensor type (
NTCT
,
SPRTD
, or
USER
) and press
ENTER
.
4.
Use the ◄ and ►keys to display an RTD parameter and press
ENTER
.
5.
Use the ◄, ►, ▲, and ▼ keys to enter the coefficient value and press
ENTER
.
6. Repeat steps 4 and 5 to set the coefficients for all the RTD parameters.
Remote operation
Commands that use the
<n>
parameter can use the
MINimum
and
MAXimum
name parameters.
MIN
selects the minimum value and
MAX
selects the maximum value.
Use the following commands to select the RTD sensor type and set the measurement units:
TEMPerature:RTD:TYPE <name>
Select RTD sensor type
<name>
= PT100, D100, F100, PT385, PT3916, USER, SPRTD, or NTCT
UNIT <name>
Set temperature measurement units
<name>
= Cel
Celsius
= Far
Fahrenheit
= K
Kelvin