7 Troubleshooting
Error code
Problem
Solution
* E09013 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E28205
Calibration missing edges
1. Motion blocked externally.
2. See additional errors:
* E09013 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E28206
Calibration extra edges
1. Actuator pushed externally.
2. See additional errors:
* E09013 Opto missing
* E09019 Resistive sensor missing.
E28207
Calibration invalid edge
Software error. Take logs and boot the
device.
E28402
Position guard failed
Unintended rotation of the rotating
unit.
E28403
Timeout
Software error. Take logs and boot the
device.
E31202
Unexpected event
Software error. Take logs and boot the
device.
E31203
Invalid target
Software error. Take logs and boot the
device.
E31204
Unexpected opto hit
Replace horizontal 3D laser assembly.
E31205
Edge to home opto not found
See E09006.
See E31204
See E09020
See E09021
See E09022.
E31206
Motor driving error
See E09020
See E09021
See E09022.
E32202
Unexpected event
Software error. Take logs and boot the
device.
ORTHOPANTOMOGRAPH
™
OP 3D
87