7 Troubleshooting
Error code
Problem
Solution
3. Replace i3000.
E57203
ACK timeout dropped
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable.
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
E57204
Invalid sender
Software error. Take logs and boot the
device.
E57205
New before end
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable
2. Replace i3000
3. Replace i2000