7 Troubleshooting
Error code
Problem
Solution
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
E57206
Mixed messages
Software error. Take logs and boot the
device.
E57207
Eof
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable.
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
E57208
Order
Software error. Take logs and boot the
device.
E57402
Cannot re enable receiving
Software error. Take logs and boot the
device.
E57403
Extra ack
Software error. Take logs and boot the
device.
E57404
Invalid ack
Error source i3000:
1. Check connections:
ORTHOPANTOMOGRAPH
™
OP 3D
93