7 Troubleshooting
Error code
Problem
Solution
3. Replace i2000
4. Replace i4000
Error source, i4000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
2. Replace i4000
3. Replace i3000.
E57003
Received message with no
service registered
Software error. Take logs and boot the
device.
E57004
Isr invalid service
Software error. Take logs and boot the
device.
E57005
Isr frame queue full
Software error. Take logs and boot the
device.
E57006
Isr interupt enabled failed
Software error. Take logs and boot the
device.
E57007
Buffer full force reboot
Software error. Take logs and boot the
device.
E57202
ACK timeout
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable
* J3009, J2010
If fail => replace the main cable
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable
2. Replace i4000
ORTHOPANTOMOGRAPH
™
OP 3D
91