7 Troubleshooting
Error code
Problem
Solution
* E09018 Resistive sensor missing.
E27204
Missed opto
1. Motion blocked externally.
2. See additional errors:
* E09012 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E27205
Calibration missing edges
1. Motion blocked externally.
2. See additional errors:
* E09012 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E27206
Calibration extra edges
1. Actuator pushed externally.
2. See additional errors:
* E09012 Opto missing
* E09018 Resistive sensor missing
E27207
Calibration invalid edge
Software error. Take logs and boot the
device.
E27402
Position guard failed
Unintended linear position change of
the rotating unit.
E27403
Timeout
Software error. Take logs and boot the
device.
E28202
Position fail
1. Pushed/blocked externally.
2. See additional errors:
* E09013 Opto missing
* E09019 Resistive sensor missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
E28203
Unexpected opto hit
1. Actuator pushed externally.
2. See additional errors:
* E09013 Opto missing
* E09019 Resistive sensor missing.
E28204
Missed opto
1. Motion blocked externally.
2. See additional errors:
86
ORTHOPANTOMOGRAPH
™
OP 3D