7 Troubleshooting
Error code
Problem
Solution
E21202
Goto Fixed - no opto found
1. Check connections:
* J3005, J1026
If fail (open) => replace cable pivot
sensor movement cable
2. Replace opto fork
3. Replace opto ring if any of the
teeth are broken
4. Replace i3000.
E21203
Calib ABS - opto hit invalid ADC
value
See E09017 Resistive sensor missing.
E21204
Calib ABS - opto hit ADC value
change too small
See E09017 Resistive sensor missing.
E21205
Calib ABS - opto hit too early
See E09017 Resistive sensor missing.
E21206
Calib ABS verify - too large error See E09017 Resistive sensor missing.
E21207
Route - no valid route found
Software error. Take logs and boot the
device.
E21208
Goto ADC - Invalid start position Software error. Take logs and boot the
device.
E21209
Goto ADC - Opto hit too fast
See E26203 Goto ADC -Opto hit too
fast.
E21210
Goto ADC - Opto edge missing
See E26204.
See E26205.
E21211
Goto ADC - Edges end not found 1. Check the location of the resistive
position sensor.
If fail => replace sensor and
reposition it.
E21212
Goto ADC - Adc value not
changing
See E09017.
E21213
Goto ADC - Adc value changing
to wrong direction
1. Check connections:
* J2002, J1012
If fail => replace rotating high torq
motor cable
* J2001, J1013
If fail => replace rotating sensor
cable.
E22002
Route - waypoints on invalid
state
Software error. Take logs and boot the
device.
E22202
Goto Fixed - no opto found
1. Check connections:
* J3004, J1025
82
ORTHOPANTOMOGRAPH
™
OP 3D