8 Part replacement instructions
2.
Remove the screws (2 pcs) from below of the rotating unit.
3.
Pull the rotator front and back covers off.
Removing the tube head back cover
Required tools: Allen key 3mm
NOTICE!
The tube head back cover is attached with two screws from top of the
rotating unit and with one screw on the tube head inner cover.
1.
Remove or loosen the screws from right and left side of the rotating unit.
2.
Remove the screw on the tube head front cover.
3.
Remove the tube head back cover.
Removing the tube head front cover
Required tools: Allen key 3mm
108
ORTHOPANTOMOGRAPH
™
OP 3D
Содержание ORTHOPANTOMOGRAPH OP 3D
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Service manual ENGLISH 215241 rev 4 0 805 4918...
Страница 2: ......
Страница 45: ...5 Diagrams 5 Diagrams 5 1 Wiring diagram ORTHOPANTOMOGRAPH OP 3D 45...