4 Device description
4.5.3 i4000 power supply board
The i4000 power supply board is located on the carriage. See figure
Locations of circuit
boards
.
The i4000 board is one of three control boards of the OP 3D X-ray imaging system. The
main features of the i4000 power supply board are listed below.
Functionality
• Generate all main supplies to OP 3D
• UI control
• Z motor control and position sensing
• Laser motor control and position sensing
• Chin rest sensor interface
• Patient positioning lights control
• Scan motor control and positioning
• Filament control
• HV control
Indicators
LED #
Color
Default state
Function
H1
Green
ON
+3V3HV is on
H2
Green
ON
#15V is on
H3
Green
ON
+20V Triac is on
H4
Green
ON
+17V is on
H5
Green
ON
HV395 is on
H6
Green
ON
PFC power is on
H7
Green
ON
P15V is on
H8
Green
OFF
Preheat is on
H9
Green
ON
P3V3 is on
H10
Green
ON
P5V is on
H11
Green
ON
+24V is on
H12
Red
OFF
Tubefail error
H13
Green
ON
+20V DCDC
H14
Yellow
OFF
Exposure on
H15
Yellow
OFF
Power detected
LA100
ON
Mains voltage on
40