8 Part replacement instructions
10.
Put the protective plate back.
11.
Screw in and tighten the three screws of the protective plate.
12.
Align the laser lights.
8.3.4 Replacing the i4000 power supply board
CAUTION!
Make sure that the power cannot be swiched on while you carry out
this procedure.
NOTICE!
You need to remove the carriage right and left cover before this
procedure.
NOTICE!
To replace the i4000 power supply board, you first need to remove the
horizontal lasers.
Required tools: screwdriver 3mm
1.
Open the clamps of the cables on the right.
120
ORTHOPANTOMOGRAPH
™
OP 3D