6 Maintenance
7.
When the exposure starts, check that all exposure warning indicators are working:
• Exposure warning indicator on the GUI
• Yellow exposure indicator light on the Positioning panel
• Yellow exposure indicator light on the Status indicator light on top of the carriage.
• Audible exposure warning signal
• Remote exposure switch (if in use)
• Warning light (if in use)
8.
When the exposure ends, check that the indicator lights go out, exposure warning
signal stops and a program end tone is played.
9.
Check that the taken image transfers to the imaging workstation.
6.1.5 Changing the fuses
Main fuses are located at the rear of the carriage, under the back panel cover. To reach
the fuses, remove the cover first.
NOTICE!
See chapter
Fuses
for details on the correct fuses.
To release a fuse cap, push it inwards and rotate counterclockwise with a flat head
screwdriver.
Remove the fuse from the socket and replace it with a new one. Install the fuse by
pushing the fuse cap in and rotating it clockwise with a screwdriver.
The 10A and 15A fuses are physically of different sizes and have their own fuse caps. If
you need to change the fuse rating, also the fuse cap needs to be changed. The fuses
are supplied with the corresponding fuse caps.
6.1.6 Monitor calibration
It is recommended to calibrate or check a test image with the primarily used workstation
monitor. Refer to the monitor/software manuals for information on how to do this.
ORTHOPANTOMOGRAPH
™
OP 3D
59