8 Part replacement instructions
8.3.10 Replacing the lower shelf upper cover assembly
Replacing the i9000 chin rest sensor with lasers
NOTICE!
You must remove the lower shelf upper cover and the lower shelf tray
before this procedure.
Required tools: Allen key
1.
Push and then pull to disconnect the connector.
2.
Remove the screw on the board.
3.
Lift the board off.
4.
Put the new board to its place.
5.
Insert the screw to the board and tighten.
6.
Connect the connector cable to the board.
Replacing the tilt light
NOTICE!
You must remove the lower shelf upper cover before this procedure.
NOTICE!
The tilt light is a part of the i9000 chin rest sensor with lasers
sparepart.
Required tools: Allen key
134
ORTHOPANTOMOGRAPH
™
OP 3D
Содержание ORTHOPANTOMOGRAPH OP 3D
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Service manual ENGLISH 215241 rev 4 0 805 4918...
Страница 2: ......
Страница 45: ...5 Diagrams 5 Diagrams 5 1 Wiring diagram ORTHOPANTOMOGRAPH OP 3D 45...