4 Device description
4.1.5 Device labels
1.
Type label
2.
Sensor label
3.
THA label
4.
Collimator label
5.
Filtration label
6.
Warning label
4.1.6 Emergency stop switch
An emergency stop switch is located on the left side of the carriage.
Pressing the emergency stop switch immediately terminates the imaging and all device
movements and resets all circuit boards.
NOTICE!
An interrupted imaging process cannot be resumed. A new image needs
to be taken.
To release the emergency stop switch, rotate it clockwise.
NOTICE!
Make sure that the emergency stop switch is not pressed down when
you start the imaging process.
The Graphical User Interface (GUI) has indication if the emergency stop is active and the
device status indication lights are blue.
NOTICE!
Activating the emergency stop causes the device to restart.
16
ORTHOPANTOMOGRAPH
™
OP 3D
Содержание ORTHOPANTOMOGRAPH OP 3D
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Service manual ENGLISH 215241 rev 4 0 805 4918...
Страница 2: ......
Страница 45: ...5 Diagrams 5 Diagrams 5 1 Wiring diagram ORTHOPANTOMOGRAPH OP 3D 45...