8 Part replacement instructions
Replacing the i6000 user interface board
NOTICE!
To replace the patient positioning panel, you need to replace the whole
upper lower shelf cover.
Replace the board by following the instructions in
Removing the lower shelf upper
and lower cover
.
8.3.11 Replacing the collimator
NOTICE!
You must remove the tube head front and back covers before this
procedure.
Required tools: Allen key 3mm
1.
Detach the three cables from their connectors from the top of the collimator.
ORTHOPANTOMOGRAPH
™
OP 3D
137