7 Troubleshooting
Error code
Problem
Solution
4. Replace the sensor.
E62021
UNEXPECTED PANEL BOOT
1. Check H9 and H10 LED on i2000
board during preparation (should be
green).
If fail => replace i2000
2. Check connection J2013 on i2000
and J0101 on PAN/3D COMBO sensor
side.
If fail => replace cable PAN 3D power
213859
3. Check connection J2015 on i2000
and PAN/3D COMBO sensor side
connection.
If fail => replace data cable PAN 3D
4. Replace sensor.
E62022
PANEL ROI SETUP ERROR
1. Check connection J2015 on i2000
and PAN/3D COMBO sensor side
connection.
If fail => replace data cable PAN 3D
2. Replace sensor.
E62023
FRONT BUFFER FAILED
1. Re-program i2000
If fail => replace i2000.
E62024
MEMORY READ FAILED
1. Re-program i2000
If fail => replace i2000.
E63002
Initialization failed.
1. Restart the device.
If still fails => replace i4000.
E63003
Communication failed
1. Restart the device.
If still fails => replace i4000.
E63402
Invalid Adc Value
1. Mains voltage dropped under
specified value on the wall socket.
2. Restart the device.
If still fails and mains voltage is in
limits => replace i4000.
E63403
Poweroff detected
1. If normal device power down =>
N/A
2. Restart the device.
If still fails and mains voltage is
in limits => check i4100 and its
connections.
if fail => replace i4100
=> check i4000 and its connections.