7 Troubleshooting
Error code
Problem
Solution
If fail (open) => Replace rotating
sensor cable
2. Replace opto fork
3. Replace belt wheel if any of teeth
are broken
4. Replace i2000.
E23203
Calib ABS - opto hit invalid ADC
value
See E09019
E23204
Calib ABS - opto hit ADC value
change too small
See E09019
E23205
Calib ABS - opto hit too early
See E09019
E23206
Calib ABS verify - too large error See E09019
E23207
Route - no valid route found
Software error. Take logs and boot the
device.
E23208
Goto ADC - Invalid start position Software error. Take logs and boot the
device.
E23209
Goto ADC - Opto hit too fast
See E28203
E23210
Goto ADC - Opto edge missing
See E28204
See E28205
E23211
Goto ADC - Edges end not found 1. Check the location of the resistive
position sensor.
If fail => Replace sensor and
reposition it.
E23212
Goto ADC - Adc value not
changing
See E09019
E23213
Goto ADC - Adc value changing
to wrong direction
1. Check connections:
* J2002, J1012
If fail => Replace rotating high torq
motor cable
* J2001, J1013
If fail => Replace rotating sensor
cable.
E26202
Position fail
1. Pushed/blocked externally.
2. See additional errors:
* E09011 Opto missing
* E09017 Resistive sensor missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
84