7 Troubleshooting
Error code
Problem
Solution
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
E10206
Calibration feedback constant
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
E10207
Calibration feedback not raising
enough
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J 7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
E10208
Calibration invalid pwm
Software error. Take logs and boot the
device.
E10209
Calibration feedback invalid on
scan
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.