7 Troubleshooting
Error code
Problem
Solution
if fail => Replace i4000.
E64002
Init failed
Software error. Take logs and boot the
device.
E64003
Start failed
Software error. Take logs and boot the
device.
E64202
Control loop too slow
Software error. Take logs and boot the
device.
E64203
Watchdog fail
Software error. Take logs and boot the
device.
E90002
No error
Software error. Take logs and boot the
device.
E90003
Invalid command
Software error. Take logs and boot the
device.
E90004
Invalid arguments
Software error. Take logs and boot the
device.
E90005
Resource not available
Software error. Take logs and boot the
device.
E90006
Error while executing
Software error. Take logs and boot the
device.
E91202
Network speed too low
1. Check the Ethernet link partner
(MWS) switch or PC settings support
Gigabit Ethernet.
2. Check that the Ethernet cabling is
CAT5e or CAT6.
3. Replace i3000.
E92202
Phy register write error
1. Restart the device.
If fail => replace i2000.
E99602
Debug point 0
Software error. Take logs and boot the
device.
E99603
Debug point 1
Software error. Take logs and boot the
device.
E99604
Debug point 2
Software error. Take logs and boot the
device.
E99605
Debug point 3
Software error. Take logs and boot the
device.
E99606
Debug point 4
Software error. Take logs and boot the
device.
98
ORTHOPANTOMOGRAPH
™
OP 3D
Содержание ORTHOPANTOMOGRAPH OP 3D
Страница 1: ...ORTHOPANTOMOGRAPH OP 3D Service manual ENGLISH 215241 rev 4 0 805 4918...
Страница 2: ......
Страница 45: ...5 Diagrams 5 Diagrams 5 1 Wiring diagram ORTHOPANTOMOGRAPH OP 3D 45...