8 Part replacement instructions
13.
Place the finger guard on top of the cable coil.
14.
Push the metal plates that hold cable in place onto the cable.
15.
Lift the rotating unit back to the grooves in the upper shelf.
NOTICE!
A clicking sound tells that the rotating unit is installed correctly to
its place.
NOTICE!
Check that the cable does not get squeezed between the rotating
unit and the upper shelf.
16.
NOTICE!
Check the marking on the cable.
Insert the main cable to the clamp.
17.
Insert the screws of the rotating unit and tighten them.
18.
Connect all the cables and connectors.
8.3.7 Replacing the i2000 image capture board
NOTICE!
You must remove the rotator back and front covers before this
procedure.
NOTICE!
Take a picture of the connectors of the i2000 image capture board
before you remove any of them.This helps you to connect the new board.
Required tools: Allen key 3mm
128
ORTHOPANTOMOGRAPH
™
OP 3D
Содержание ORTHOPANTOMOGRAPH OP 3D