Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.1 Control unit service program
MP: 2
commis‐
sioning /
adjust‐
ments
Menu for starting up the laser, calibration, et ce‐
tera.
MP 2-0: select energy power meter
Select the connected energy meter
MP 2-1: resonator adjustment
Laser pulse generator with 6Hz/80mJ for adjus‐
ting the resonator
MP 2-2: deflection mirror adjustment
Laser pulse generator with 1Hz/60mJ for adjus‐
ting the deflection mirror
MP 2-3: laser hose coupling adjustment
Laser pulse generator with 15Hz/250mJ for ad‐
justing the hose input
MP 2-4: power measuring unit adj. (VEM)
Adjustment of the energy measuring electronics
(VEM)
MP 2-5: automatic energy adjustment
Automatic adjustment of the energy using an ex‐
ternal meter
MP 2-6: manual energy adjustment
Manual, stepwise energy adjustment (semi au‐
tomatic)
MP 2-7: manual laser pulse generator
Manual laser pulse generator for testing or ad‐
justing an individual frequency/energy combina‐
tion
MP 2-8: display control tables
Display of the control and measured value tables
MP 2-9: energy safety check
Safety check of energy
MP: 3 sta‐
tic data
setup
Menu for software options/initialisation
MP 3-0: standard initialisation
Standard initialisation
MP 3-1: change date
Change date
MP 3-2: change time
Change time
MP 3-3: choose language
Choose a language (depending on the available
language files)
MP 3-4: software setup
Not provided for a service; do not change
MP 3-5: factory - commissioning mode
On/off of KaVo Factory startup mode
MP 3-6: send user.dat
Send user.dat to service computer
MP 3-7: reload EEP-file to display
Sends k3_sde.eep to re-initialise the display unit
MP 3-8: ON/OFF optimised feedback
Only for clinical tests - has to be turned off
MP 3-9: detect/feedback/US en-/disable
detect/feedback/US code
MP: 4
Diagno‐
dent Ser‐
vice
Menu for Diagnodent functions
MP 4-0: unmodulated pilot laser output
Unmodulated operation of the pilot laser for mea‐
suring output
MP 4-1: modulated pilot laser output
Diagnodent test point with measured value out‐
put
MP 4-2: pilotlaser current measurement
Measurement of the current through the pilot la‐
ser diode
MP 4-3: direct diagnodent output
Not intended for service
MP 4-4: operating period
Display of the Diagnodent operating hours
MP 4-5: clear operating period
Reset operating hours
MP: 5
Start wi‐
thout se‐
curity
system
Start the operating mode without the safety shu‐
toffs
Only for test purposes for the service technician
MP: 6
Software
download
Receive mode for file transfer via service inter‐
face - service computer
(ZMODEM protocol, 19200 baud)
MP: 7
Power off
Turnoff device, software shutdown
25/176