Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.1 Control unit service program
2. Control.exe
3. Service.exe
4. Ztrans.exe
5. Default.dat
6. Default.usr
7. *.stx(gr.stx)
8. *.prg(gr.prg)
9. k3_sde.eep
10. Rominit.bat
*The number of these files can vary depending on the number of languages in user
mode. Observe the information in the service news on the software update.
▶ Select the Zmodem protocol.
▶ Execute the Send command in the communications program
The touch screen display shows the transfer status with file check and, once
transfer has been completed successfully:
Transfer end.
To return to the service mode main menu, choose Continue.
Note
Always send all of the files of a software update to the KEY Laser. All of the files
in the KEY Laser may only correspond to a single software version.
2.1.14 User mode error messages/causes
Notes/warnings
Meaning
FOOTSWITCH HELD DOWN ALREADY!
The footswitch button has been depressed.
FIBRE NOT CONNECTED!
Fibre (laser hose) contact not connected to laser hose
coupling
SPRAY CONTAINMENT EMPTY
Spray containment empty
DOOR OPEN! (DOOR INTERLOCK)
Door contact open/No strapping plug
INVALID PROGRAM NAME!
Program name already used
NO MORE PROGRAMS CAN BE CREATED!
Program memory full/max. number of user-defined
programs reached.
HOLDER SWITCH ACTIVATED!
Swing arm already activated at switch-on
HOLDER SWITCH NOT ACTIVATED!
At laser pulse activation, shutter circuit open/swing
arm contact open/footswitch contact open
INTERNAL COOLING WATER TEMPERATURE TOO
HIGH
Cooling water temperature too high/analogue Temp
temperature signal has exceeded the maximum per‐
missible limit -> cooling mode is active.
Error message
Explanation/possible cause errors
SYSTEM ERROR 1
Error with I²C access to the VEM/Line interruption
SYSTEM ERROR 2
Internal software error, incorrect IMR for standby cir‐
cuit
SYSTEM ERROR 3
Cannot load the standard program file (*.prg)
SYSTEM ERROR 4
Error saving the counter in EEPROM, EEPROM struc‐
ture destroyed, reset the counter
71/176