![KaVo KEY Laser III 1243 Скачать руководство пользователя страница 127](http://html.mh-extra.com/html/kavo/key-laser-iii-1243/key-laser-iii-1243_technicians-instructions_658195127.webp)
Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.2 Optical system
2.2.6 Measure laser hose coupling
① Er:YAG crystal
③ Replacement O-rings
② Input block
Note
The lens in the input block is harmonised with the Er:YAG crystal in the Z axis.
This is done by a spacing ring that is made by KaVo specially for this adjustment.
The spacer ring is mounted in the lens holder of the input block.
The input block of the lens of the Er:YAG crystal can therefore only be exchanged
by exchanging the entire assembly: Measure exchange input block
Mat. no. 1.001.7692Mat. no. 1.001. 7691 (new).
The "Measure input block" kit contains an Er:YAG crystal 1 and an input block 2
that is adjusted to the provided Er:YAG crystal in the Z axis. Two replacement O-
rings for the Er:YAG crystal are also included.
125/176